Abstract
We have investigated using a water and tetraethylene glycol mixture to reduce, substantially, the equilibrium vapor pressure of water in our expansion cloud chamber. The experiments were performed with a mixture that reduced the water vapor content to 2.5% relative humidity. They verify that the chamber with this mixture can be expanded to temperatures from -30° C to -40° C accompanied by water vapor pressures from below saturation to supersaturation ratios up to 1.6. The experiments indicate that the very low concentration of tetraethylene glycol in the vapor does not influence the water nucleation process.