Abstract
We propose a novel method to overcome significant problems of baseline drift and sensitivity degradation in amperometric biosensors based on oxidase enzyme reactions. A novel glucose microsensor with a built-in electrochemical oxygen manipulation microsystem is introduced to demonstrate three novel functionalities; one-point in situ self-calibration (zero-point), broadening of dynamic range and increase in sensitivity. The influence of electrochemically generated oxygen microenvironment on the sensor output within a fluidic structure is investigated.
| Original language | American English |
|---|---|
| Journal | Proceedings of the 3rd IEEE/EMBS Special Topic Conference on Microtechnology in Medicine and Biology, 2005 |
| DOIs | |
| State | Published - Jan 1 2005 |